Sizes of these terms reflect their relevance to your search.

Total hip replacement (THR) is replacing the prosthesis stem similar to human bone that takes advantage of the material with both good mechanical properties and biocompatibility to the damaged articular surface. Thus it can not only alleviate or even eliminate the pain but also effectively maintain the joint stability and freedom and restore its normal performance. Finite element analysis was used in this study to establish a 3D model of artificial hip stem, and explore its fatigue properties of different materials to ensure the safety and reliability. The calculating obtained two results of different metal hip prosthesis, including lifetime and deformation. The minimum service life of titanium prosthesis reaches 568 million times, which satisfies ISO standards, while the stainless steel does not suit to be a prosthesis material.

Citation

Gang Tang, Jiange Wang, Hongxia Luo. Fatigue property analysis of prosthesis of hip joint with two different materials]. Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi. 2015 Feb;32(1):73-6

Expand section icon Mesh Tags

Expand section icon Substances


PMID: 25997269

View Full Text